Is the zero point calibration of the measuring modules carried out to an average value or to the current instantaneous value?
With zero adjustment via CAN message, the currently applied measured value is used (no mean value or similar.). The user must observe the system and ensure that the input measurement values do not oscillate and that the system behaves "quietly". Noise can falsify the zeroing.