The STG CAM MM does not perform a remote bridge adjustment to the specified CAN ID?
Unlike the ADMM CAN, the STGMM CAN does not react to a message with any content. The CAN message, which is to trigger the calculation of the zero offset for the respective channel in the measurement module, must have the following content:
Module type | DLC | Byte 0 | Byte 1 | Byte 2 | Byte 3 | Byte 4 | Byte 5 | Byte 6 | Byte 7 |
---|---|---|---|---|---|---|---|---|---|
STG CAN MM | 8 | 0 | 0 | 0 | 0 | 0x7A | 0x65 | 0x72 | 0x6F |